If you want to measure the degree of a liquid easily and reliably, a lot of people can do this using hydrostatic pressure measurement, e.g. with a submersible pressure transmitter or perhaps a so called level probe. The characteristic submersed application implicates a maximum exposure to the surrounding, mainly water-based medium, respectively to ?moisture?.
Exposure isn’t only limited by the wetted elements of the pressure sensor housing, but also to the complete immersed amount of the cable. Furthermore, outside Unassuming immersed level probe parts, the cable, and in particular the cable end, are often exposed to moisture as a result of splash water, rain and condensation. This is true not merely during operation, but a lot more during installation and commissioning, or when maintenance or retrofitting is required. Irrespective of the target application, whether in water and wastewater treatment or in tank monitoring, moisture ingress in to the cable ends of the submersible pressure transmitter can occur early and irreversibly with insufficient protection measures, and, in almost all cases, result in premature failure of the instrument.
The ingress of moisture into the cable outlet and from there on downwards in to the electronics of the level probe must be actively eliminated by preventive actions by the user. To measure the level with highest accuracy, the varying ambient pressure above the liquid media, that is also ?resting? on the liquid, should be compensated contrary to the hydrostatic pressure acting on the pressure sensor (see article: hydrostatic level measurement).
Ventilation tube
Thus, it is logical that there surely is a constant threat of a moisture-related failure because of moisture ingress (both via the ventilation tube and through the specific cable itself) if you can find no adequate protective measures. To pay the ambient pressure ?resting? on the media, a ventilation tube runs from the sensor element within the level probe, through the cable and out of the level probe by the end of the cable. Due to capillary action within the ventialation tube used for ambient pressure compensation, moisture may also be transported from the surrounding ambience right down to the sensor.
Thus not merely air, but additionally moisture penetrates in to the tube, hence the sensor in the probe and the electronics around it is usually irreparably damaged. This can lead to measurement errors and, in the worst case, even to failure of the level probe. To avoid any premature failure, the ingress of moisture in to the ventilation tube should be completely prevented. Additional protection against moisture penetration through the ventilation tube is supplied by fitting an air-permeable, but water-impermeable filter element by the end of the vent tube.
bare wires
Not to be ignored can be the transport of the liquid through high-humidity loads along the only limitedly protected internals of the cable, e.g. along the wires, completely right down to the submersible pressure transmitter. As a respected manufacturer, WIKA uses appropriate structural design to prevent fluid transport, as far as possible, in to the electronics of the submersible pressure transmitter. Because of molecular diffusion and capillary effects, a guaranteed one-hundred percent protection on the full duration of the submersible pressure transmitter, however, is never achievable.
It is therefore recommended that the cable is definitely terminated in a waterproof junction box with the appropriate IP protection (e.g. IP65) which is matched to the installation location. If this cable junction box is subjected to weather and varying temperature conditions, it is also recommended to pay attention to a controlled pressure equalisation to be able to avoid the formation of condensation or perspiration water and pumping effects. To address this technical requirement, as an accessory to a submersible pressure transmitter, you’ll be able to order a link box with an integrated air-permeable, water-impermeable membrane.
Ultimately, moisture ingress can occur not only through the exposed end of the cable, but additionally through mechanical damage to the cable sheath or due to liquid diffusion due to improper chemical resistance of the cable material. In the article ?Selection criteria for the prevention of moisture-related failures of submersible pressure transmitters or level probes? this failure mode is described at length.
WIKA offers comprehensive solutions for the hydrostatic-pressure level measurement. For further assistance in selecting the submersible pressure transmitter most suitable for the application, please use our contact form.
Please find more info on this topic on our information platform ?Hydrostatic level measurement?

Leave a Reply