Preventing the ingress of moisture through the cable of submersible pressure transmitters and level probes?

If you would like to measure the degree of a liquid easily and reliably, a lot of people will do this using hydrostatic pressure measurement, e.g. with a submersible pressure transmitter or perhaps a so called level probe. The characteristic submersed application implicates a maximum contact with the encompassing, mainly water-based medium, respectively to ?moisture?.
Exposure is not only limited by the wetted parts of the pressure sensor housing, but also to the entire immersed length of the cable. Furthermore, beyond your directly immersed level probe parts, the cable, and specifically the cable end, are often exposed to moisture due to splash water, rain and condensation. That is true not merely during operation, but a lot more during installation and commissioning, or when maintenance or retrofitting is necessary. Irrespective of the target application, whether in water and wastewater treatment or in tank monitoring, moisture ingress into the cable ends of the submersible pressure transmitter can occur early and irreversibly with insufficient protection measures, and, in almost all cases, lead to premature failure of the instrument.
The ingress of moisture into the cable outlet and from there on downwards in to the electronics of the level probe should be actively eliminated by preventive actions by the user. To gauge the level with highest accuracy, the varying ambient pressure above the liquid media, that is also ?resting? on the liquid, should be compensated contrary to the hydrostatic pressure functioning on the pressure sensor (see article: hydrostatic level measurement).
Ventilation tube
Thus, it is logical that there is a constant threat of a moisture-related failure because of moisture ingress (both via the ventilation tube and through the specific cable itself) if you can find no adequate precautionary measures. To compensate the ambient pressure ?resting? on the media, a ventilation tube runs from the sensor element within the level probe, through the cable and from the level probe by the end of the cable. Due to capillary action within the ventialation tube used for ambient pressure compensation, moisture may also be transported from the encompassing ambience right down to the sensor.
Thus not only air, but also moisture penetrates into the tube, hence the sensor inside the probe and the electronics around it really is irreparably damaged. This may lead to measurement errors and, in the worst case, even to failure of the particular level probe. To prevent any premature failure, the ingress of moisture into the ventilation tube must be completely prevented. Challenge against moisture penetration through the ventilation tube is supplied by fitting an air-permeable, but water-impermeable filter element at the end of the vent tube.
bare wires
Not to be ignored can be the transport of the liquid through high-humidity loads across the only limitedly protected internals of the cable, e.g. along the wires, all the way down to the submersible pressure transmitter. As a leading manufacturer, WIKA uses appropriate structural design to prevent fluid transport, as far as possible, into the electronics of the submersible pressure transmitter. Because of molecular diffusion and capillary effects, a guaranteed one-hundred percent protection over the full duration of the submersible pressure transmitter, however, is never achievable.
It is therefore recommended that the cable is definitely terminated in a waterproof junction box with the correct IP protection (e.g. IP65) which is matched to the installation location. If this cable junction box is subjected to weather and varying temperature conditions, it is also recommended to pay attention to a controlled pressure equalisation in order to prevent the formation of condensation or perspiration water and pumping effects. To address this technical requirement, being an accessory to a submersible pressure transmitter, you’ll be able to order a connection box with an integrated air-permeable, water-impermeable membrane.
Ultimately, moisture ingress can happen not only through the exposed end of the cable, but additionally through mechanical harm to the cable sheath or due to liquid diffusion due to improper chemical resistance of the cable material. In this article ?Selection criteria for preventing moisture-related failures of submersible pressure transmitters or level probes? this failure mode is described at length.
WIKA offers comprehensive solutions for the hydrostatic-pressure level measurement. For further assistance in selecting the submersible pressure transmitter the most suitable for the application, please use our contact page.
Please find more info on this topic on our information platform ?Hydrostatic level measurement?